I tell the residents: "The pacemaker is just common sense: if there is no beat, it provides one; if there is one, it keeps itself from pacing."
This is similar to Ken Grauer's comment at the bottom:
"What would I do if I were a pacemaker?"
DDD pacer = Dual paced, Dual sensed, Dual response (triggered and inhibited)
Dual means that this occurs in both the atrium and the ventricle.
As you can see from the chart below, a VVI pacer is ventricle only and cannot be triggered by an atrial beat. It can only be inhibited by an intrinsic ventricular beat.
So a DDD pacer can pace the atrium (if needed), then wait a defined period (e.g., 200 ms, like a PR interval) and pace the ventricle (if needed). Either or both functions can be inhibited if the atrium or ventricle (or both) fires before the interval passes.
One of the reasons I'm presenting this case is that several residents misunderstood the concept of "failure to sense," believing it to be a cause of absence of pacing. This made me realize that pacemaker function is not as well understood as I thought.
However, failure to sense is NOT a reason for absence of pacing. Failure to sense results in inappropriate pacing: when the pacer is not sensing intrinsic activity that is present (failure to sense), it fails to have its pacing appropriately inhibited. (If a heart beat comes soon enough, the pacer should sense it and inhibit pacing so that there is no pacing when it's not needed.)
In fact, when a pacemaker fails to generate an impulse (fails to pace), one reason may be oversensing, not undersensing.
This is why we use a magnet when there is failure to pace! To stop oversensing and thus stop inhibtion of pacing. The magnet turns off the sensing function and the pacer becomes "asynchronous," meaning it will fire at a regular rate regardless of the intrinsic activity of the heart. We place the magnet because a failure to pace may be due to oversensing (not undersensing) with inappropriate inhibition of pacing.
Case
A patient called 911 for weakness.
She had recently had a DDD pacemaker placed.
Her prehospital ECG showed failure to pace by the internal pacer. External pacing was attempted but not tolerated and, because the patient was only mildly hypoperfused, she was transported to the ED without further intervention.
Exam revealed a fresh surgical scar at the area of pacemaker placement with significant swelling due to hematoma at the site of insertion.
Here is the first ED ECG:
So what is happening here? Failure to capture. There may also be failure to sense, but it is not having any effect.
I'll explain with the arrows:
Clinical course:
The potassium was normal, there was no ischemia or drug toxicity.
She went for emergent pacemaker revision. It turned out that the hematoma at the "pocket" of the pacemaker had displaced the pacemaker, pulling the ventricular lead out of position. This was fixed.
Etiologies of Failure to Capture
This was a failure to pace due to Failure to Capture (not due to oversensing, as there were pacer spikes), the differential diagnosis of which is:
--Hyperkalemia
--Ischemia
--Drug toxicity
--Poor electrode contact (as in this case)
--Fibrosis around electrode tip
Failure to Pace also includes these etiologies, none of which will produce a pacing spike on the surface ECG:
--Oversensing
--Broken leads (assess by Chest X-ray)
--Leads detached from generator
--Generator malfunction, including battery depletion and programming error
Emergency Management:
1. Treat hyperkalemia or drug toxicity
2. Treat ischemia (Cath lab)
3. Magnet application if no pacer spikes.
4. External pacing
5. Transvenous pacing
6. Chronotropic support with Isoproterenol, Epinephrine, or Dopamine to increase ventricular escape rate.
____________
John Larkin, of the ECG of the week (http://jhcedecg.blogspot.com/), has this simple thought process:
Hi Steve,
Great case as always, I just thought I'd share my very simple interpretation of some of the issues seen with PPMs.
Sensing is about how many spikes are there and how many should there be.
Capture is about what is the effect of the spike.
Regarding under vs over sensing I use a very simple aide memoire of:
Over-sensing = Under pacing (less pacing spike than you'd anticipate, due to excessive inhibition)
Under-sensing = Over pacing (more spike that you'd anticipate, reduced / lack of inhibition from native cardiac activity).
John L
This is similar to Ken Grauer's comment at the bottom:
"What would I do if I were a pacemaker?"
DDD pacer = Dual paced, Dual sensed, Dual response (triggered and inhibited)
Dual means that this occurs in both the atrium and the ventricle.
As you can see from the chart below, a VVI pacer is ventricle only and cannot be triggered by an atrial beat. It can only be inhibited by an intrinsic ventricular beat.
So a DDD pacer can pace the atrium (if needed), then wait a defined period (e.g., 200 ms, like a PR interval) and pace the ventricle (if needed). Either or both functions can be inhibited if the atrium or ventricle (or both) fires before the interval passes.
One of the reasons I'm presenting this case is that several residents misunderstood the concept of "failure to sense," believing it to be a cause of absence of pacing. This made me realize that pacemaker function is not as well understood as I thought.
However, failure to sense is NOT a reason for absence of pacing. Failure to sense results in inappropriate pacing: when the pacer is not sensing intrinsic activity that is present (failure to sense), it fails to have its pacing appropriately inhibited. (If a heart beat comes soon enough, the pacer should sense it and inhibit pacing so that there is no pacing when it's not needed.)
In fact, when a pacemaker fails to generate an impulse (fails to pace), one reason may be oversensing, not undersensing.
This is why we use a magnet when there is failure to pace! To stop oversensing and thus stop inhibtion of pacing. The magnet turns off the sensing function and the pacer becomes "asynchronous," meaning it will fire at a regular rate regardless of the intrinsic activity of the heart. We place the magnet because a failure to pace may be due to oversensing (not undersensing) with inappropriate inhibition of pacing.
Case
A patient called 911 for weakness.
She had recently had a DDD pacemaker placed.
Her prehospital ECG showed failure to pace by the internal pacer. External pacing was attempted but not tolerated and, because the patient was only mildly hypoperfused, she was transported to the ED without further intervention.
Exam revealed a fresh surgical scar at the area of pacemaker placement with significant swelling due to hematoma at the site of insertion.
Here is the first ED ECG:
What are you seeing? Try to figure it out before reading the explanation below. |
So what is happening here? Failure to capture. There may also be failure to sense, but it is not having any effect.
I'll explain with the arrows:
Clinical course:
The potassium was normal, there was no ischemia or drug toxicity.
She went for emergent pacemaker revision. It turned out that the hematoma at the "pocket" of the pacemaker had displaced the pacemaker, pulling the ventricular lead out of position. This was fixed.
Etiologies of Failure to Capture
This was a failure to pace due to Failure to Capture (not due to oversensing, as there were pacer spikes), the differential diagnosis of which is:
--Hyperkalemia
--Ischemia
--Drug toxicity
--Poor electrode contact (as in this case)
--Fibrosis around electrode tip
Failure to Pace also includes these etiologies, none of which will produce a pacing spike on the surface ECG:
--Oversensing
--Broken leads (assess by Chest X-ray)
--Leads detached from generator
--Generator malfunction, including battery depletion and programming error
Emergency Management:
1. Treat hyperkalemia or drug toxicity
2. Treat ischemia (Cath lab)
3. Magnet application if no pacer spikes.
4. External pacing
5. Transvenous pacing
6. Chronotropic support with Isoproterenol, Epinephrine, or Dopamine to increase ventricular escape rate.
____________
John Larkin, of the ECG of the week (http://jhcedecg.blogspot.com/), has this simple thought process:
Hi Steve,
Great case as always, I just thought I'd share my very simple interpretation of some of the issues seen with PPMs.
Sensing is about how many spikes are there and how many should there be.
Capture is about what is the effect of the spike.
Regarding under vs over sensing I use a very simple aide memoire of:
Over-sensing = Under pacing (less pacing spike than you'd anticipate, due to excessive inhibition)
Under-sensing = Over pacing (more spike that you'd anticipate, reduced / lack of inhibition from native cardiac activity).
John L
=====================================
My Comment, by KEN GRAUER, MD (11/9/2018):
=====================================
Interpretation of pacemaker tracings used to be straightforward and surprisingly simple when all the provider had to worry about were VVI pacemakers. These devices paced the ventricles — sensed the ventricles — and were inhibited by a QRS complex. Systems are far more complex these days, with dual chamber pacemakers and numerous pacing schemes, including less intuitive rate-responsive functions. As a result, determining whether a pacemaker is appropriately functioning or not has become exceedingly challenging — especially when parameters that the pacer has been set to are not known. Two thoughts have always helped me when confronted with a problematic pacemaker tracing:- FIRST: Trying to imagine,“What would I do if I was the pacemaker” — with my sole function in life being to respond at a certain heart rate if no spontaneous impulse “from above” is forthcoming — and/or — inhibit myself if a response from above does arrive within an appropriate time period; and,
- NEXT: Remembering that most of the time, these amazing devices (pacemakers) are appropriately performing what they have been programmed to do — so our task as ECG interpreters is often best served by assuming the pacemaker is probably correct, and trying to figure out WHY the pacer is acting as it is.
The above said — this case clearly is one in which the pacemaker is not functioning correctly. As is superbly illustrated and explained by Dr. Smith — the pacemaker IS sensing spontaneous atrial activity — and IS appropriately putting out pacer spikes — but for the most part, is not capturing the ventricles. There is no way this pacemaker could be functioning appropriately with an effective ventricular rate in the 20s!
excellent. thank you, Pendell
ReplyDeletetom, this was actually my post.
DeleteSteve
Is step 3. Magnet application with no pacer spikes something appropriate to do in the pre-hospital scope of practice (EMt-Intermediate or Paramedic level)? I'd think we would go to the eternal pacing without worrying about the magnet.
ReplyDeleteI think if you see no pacer spikes, the magnet is far easier than pacing and should be tried first
DeleteHi Steve,
ReplyDeleteGreat case as always, I just thought I'd share my very simple interpretation of some of the issues seen with PPMs.
Sensing is about how many spikes are there and how many should there be.
Capture is about what is the effect of the spike.
Regarding under vs over sensing I use a very simple aide memoire of:
Over-sensing = Under pacing (less pacing spike than you'd anticipate, due to excessive inhibition)
Under-sensing = Over pacing (more spike that you'd anticipate, reduced / lack of inhibition from native cardiac activity).
John L